An instrument that measures the characteristic lifetime of fluorescence of chlorophyll has been invented for in situ, real-time oceanographic studies of photosynthesis in phytoplankton. The basic design and principle of operation lend themselves to development of the instrument as a relatively inexpensive, sensitive, compact, rugged, portable, low-power-consumption, hand-held, shipboard unit. Similar units with designs adapted to agricultural applications (e.g., assessment of physiological statuses of crops) are also envisioned.
The need for this or a similar instrument arises because fluorescence
lifetimes are robust measures of physical and chemical mechanisms that affect
photosynthesis and of the photosynthetic productivity of phytoplankton. The
fluorescence lifetimes of particular interest are those associated with fluorescent
de-excitation in photosystem II. ("Photosystem II" denotes a series of photosynthetic
reactions in which a pigment absorbs light at wavelengths up to 680 nm and absorbed
light energy causes the splitting of water molecules, giving rise to oxygen
and to a high-energy reductant.) These lifetimes range from about 2 ns down
to as little as hundreds of picoseconds. Therefore, in order to exploit the
full potential of fluorescence-lifetime measurements for detecting small changes
in photosystem-II physiological status, this or any similar instrument would
have to exhibit an error <50 ps and a resolution
10 ps.
The present instrument does not measure fluorescence lifetimes
directly; instead, it is based on a principle of phase fluorometry, which can
be implemented more easily. In phase fluorometry, a sample is excited with light
modulated sinusoidally at an angular frequency w. The resulting fluorescence
emitted by the sample is modulated at the same frequency but, because of the
finite lifetime of the excited state, is delayed in phase by an angle f
relative to the excitation. The phase angle Ø
is measured and used to calculate a phase lifetime (
p)
according to
p
= tan Ø/w.
In order to be able to determine tp accurately by this method, one
must choose a modulation frequency comparable to the rate of decay of the fluorescence.
In the instrument (see figure), a baseband signal of frequency f0 (typically 600 Hz) is generated for use both as a sample signal and as a reference signal for the phase measurement. Using a single-sideband technique, the sample signal is up-converted by combining it with a carrier signal of frequency f (typically 70 MHz). The up-converted signal (at frequency f0 + f) is used to modulate a light-emitting diode (typical wavelength 670 nm), the output of which is used to excite the sample.
The fluorescence emitted by the sample acquires a phase delay
that corresponds to a frequency-dependent weighted average of the lifetimes
of the fluorophores present in the sample. The fluorescence, which is modulated
at frequency f0 + f, is detected by a photomultiplier
tube. The electrical output of the photomultiplier is down-converted to f0
by mixing it with the carrier signal and low-pass filtering the product signal.
The phase information acquired by the up-converted signal through interaction
with the sample is preserved in the down-converted signal. The phase-shifted
down-converted f0 sample signal and the reference f0
signal are captured by an analog-to-digital converter, and the output of the
analog-to-digital converter is stored and processed to determine Ø
and
p.
This Instrument Measures the Phase Difference between
sinusoidally modulated optical excitation and the resulting fluorescence. The
fluorescence lifetime is then calculated from the phase difference and the modulation
frequency.
This work was done by Salvador M. Fernandez, Ernest F. Guignon, Robert Kersten, and Ernest St. Louis of Ciencia, Inc., for Stennis Space Center. For further information, access the Technical Support Package (TSP) free on-line at www.nasatech.com under the Physical Sciences category.
In accordance with Public Law 96-517, the contractor has elected to retain title to this invention. Inquiries concerning rights for its commercial use should be addressed to
Ciencia, Inc.
111 Roberts Street
Suite K
East Hartford, CT 06108
Refer to SSC-00110, volume and number of this NASA Tech Briefs issue, and the page number.
Download detailed Technical Support Package for this Brief
Dedicated to helping you design better products in a digital world... your guide to the latest tools & techniques for digital prototyping, simulation, and analysis of the real-world performance of your ideas. Visit the Digital Design Center